CVE-2025-54304 Details
Description
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
A vulnerability allowing unauthorized remote code execution has been identified in the Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When powered on, these devices start an X11 display server that listens on all network interfaces via port 6000. The default X11 access control allows connections from localhost and a specific IP address. However, if the device is connected to a DHCP network and does not receive the designated IP, the display server becomes accessible to other devices on the network. Exploitation of this vulnerability involves interacting with the exposed X11 server, which can lead to gaining root privileges and executing code remotely by manipulating the matchbox-desktop environment to spawn a terminal. This issue affects a discontinued product that is no longer supported by the manufacturer.
Users are advised to verify that the Ion OneTouch 2 device is not using default login credentials and to change them if necessary. Additionally, review the network configuration to minimize exposure to security vulnerabilities. Ensure the device is not connected to the public Internet without proper security measures, such as a firewall, to prevent unauthorized access.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
CISA-ADP
Assessed Dec 5, 2025References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-200 | Exposure of Sensitive Information to an Unauthorized Actor | CISA-ADP |
Affected Products
| Product | Versions |
|---|---|
| thermofisher ion torrent onetouch 2 firmware | All versions |
CPE
Remediation
| |
| thermofisher ion torrent onetouch 2 | All versions |
CPE
Remediation
| |
Change History
6 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Nov 4, 2026 | CVE Translated | [email protected] |
| Jun 17, 2026 | CVE Modified | [email protected] |
| Jun 17, 2026 | CVE Modified | CISA-ADP |
| Dec 16, 2025 | Initial Analysis | [email protected] |
| Dec 5, 2025 | CVE Modified | CISA-ADP |
| Dec 4, 2025 | New CVE Received | [email protected] |