Not a U.S. government website. NDD is an independent vulnerability database by Volerion and is not affiliated with or endorsed by NIST or NVD.
VOLERION
Volerion Security Research

NOT DEFERRED DATABASE

VULNERABILITIES

CVE-2025-3425 Details

ANALYZED


This CVE record has been analyzed and enriched by NVDAPI.com as an independent party.

Description

The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the deserialization vulnerability. After analyzing the configuration files, we observed that the server had set the TypeFilterLevel to Full which is dangerous as it can potentially lead to remote code execution using deserialization. This issue affects IntelliSpace Portal: 12 and prior.

Metrics

CVSS 3.x Severity and Vector Strings:

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.

Weakness Enumeration

CWE-IDCWE NameSource
CWE-502Deserialization of Untrusted DataPhilips

Affected Products

ProductVersions
Philips IntelliSpace Portal
<= 12

CPE

  • cpe:2.3:a:philips:intellispace_portal:*:*:*:*:*:*:*:*

Remediation

  • Workaround:low effort

    Set the TypeFilterLevel to Medium or Low to prevent remote code execution via deserialization.