CVE-2024-48966 Details
UNDERGOING ENRICHMENT
The CVE is currently being enriched by the Volerion team. The enrichment process results in the association of reference link tags, CVSS, and CPE applicability statement data.
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Metrics
CVSS 4.0 Severity and Vector Strings:
VolerionCVSS-B:…Vector:This vulnerability is currently being analyzed.No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
VolerionCVSS-B:…Vector:This vulnerability is currently being analyzed.No data available for CVSS Version 2.0 on this CVE.
CISA-ADP
Assessed Nov 15, 2024References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01 | [email protected] |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-306 | Missing Authentication for Critical Function | [email protected] |
Affected Products
This vulnerability is currently being analyzed
Change History
3 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| Jun 17, 2026 | CVE Modified | CISA-ADP |
| Nov 14, 2024 | New CVE Received | [email protected] |