Not a U.S. government website. NDD is an independent vulnerability database by Volerion and is not affiliated with or endorsed by NIST or NVD.
VOLERION
Volerion Security Research

NOT DEFERRED DATABASE

VULNERABILITIES

CVE-2023-0888 Details

Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Metrics

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.

Weakness Enumeration

CWE-IDCWE NameSource
CWE-94Improper Control of Generation of Code ('Code Injection')[email protected]
CWE-95Improper Neutralization of Directives in Dynamically Evaluated Code ('Eval Injection')[email protected]

Affected Products

ProductVersions
bbraun battery-pack sp with wifi firmware
<= 053l000092
<= 054u000092

CPE

  • cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:global:*:*:*
  • cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:us:*:*:*

Remediation

  • No remediation found in references.
bbraun battery-pack sp with wifi
All versions

CPE

  • cpe:2.3:h:bbraun:battery-pack_sp_with_wifi:-:*:*:*:*:*:*:*

Remediation

  • No remediation found in references.

Change History

6 change records found show changes


QUICK INFO

CVE Dictionary Entry:
CVE-2023-0888
NVD Published Date:
Mar 13, 2023
NVD Last Modified:
Jun 17, 2026
Source:
[email protected]
CVE-2023-0888 Details - Not Deferred