Not a U.S. government website. NDD is an independent vulnerability database by Volerion and is not affiliated with or endorsed by NIST or NVD.
CVE-2022-43557 Details
Description
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
CISA-ADP
Assessed Apr 23, 2025Exploitation: NoneAutomatable: NoTechnical Impact: Partial
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-rs-232-interface-vulnerability | CVE | MitigationVendor Advisory |
| https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-rs-232-interface-vulnerability | [email protected] | MitigationVendor Advisory |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-287 | Improper Authentication | [email protected] |
| CWE-1299 | Missing Protection Mechanism for Alternate Hardware Interface | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| bd bodyguard 999-603 firmware | All versions |
CPE
Remediation
| |
| bd bodyguard 999-603 | All versions |
CPE
Remediation
| |
| bd bodyguard duo 999-903 firmware | All versions |
CPE
Remediation
| |
| bd bodyguard duo 999-903 | All versions |
CPE
Remediation
| |
| bd bodyguard epidural 999-683 firmware | All versions |
CPE
Remediation
| |
| bd bodyguard epidural 999-683 | All versions |
CPE
Remediation
| |
| bd bodyguard pain manager 999-803 firmware | All versions |
CPE
Remediation
| |
| bd bodyguard pain manager 999-803 | All versions |
CPE
Remediation
| |
| bd bodyguard t 999-103 firmware | All versions |
CPE
Remediation
| |
| bd bodyguard t 999-103 | All versions |
CPE
Remediation
| |
| bd bodyguard 323 colorvision firmware | All versions |
CPE
Remediation
| |
| bd bodyguard 323 colorvision | All versions |
CPE
Remediation
| |
| bd bodyguard 121 twins firmware | All versions |
CPE
Remediation
| |
| bd bodyguard 121 twins | All versions |
CPE
Remediation
| |
Change History
6 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| Jun 17, 2026 | CVE Modified | CISA-ADP |
| Nov 21, 2024 | CVE Modified | CVE |
| May 14, 2024 | CVE Modified | [email protected] |
| Oct 27, 2023 | CVE Modified | [email protected] |
| Dec 9, 2022 | Initial Analysis | [email protected] |