CVE-2020-25183 Details
Description
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
No SSVC data is available for this CVE.
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-287 | Improper Authentication | [email protected] |
| CWE-287 | Improper Authentication | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| medtronic mycarelink smart model 25000 firmware | All versions |
CPE
Remediation
| |
| medtronic mycarelink smart model 25000 | All versions |
CPE
Remediation
| |
Change History
5 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| May 22, 2025 | CVE Modified | [email protected] |
| Nov 21, 2024 | CVE Modified | CVE |
| May 14, 2024 | CVE Modified | [email protected] |
| Dec 15, 2020 | Initial Analysis | [email protected] |