CVE-2020-12040 Details
Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
No SSVC data is available for this CVE.
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | CVE | Third Party AdvisoryUS Government Resource |
| https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | [email protected] | Third Party AdvisoryUS Government Resource |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-319 | Cleartext Transmission of Sensitive Information | [email protected] |
| CWE-319 | Cleartext Transmission of Sensitive Information | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| baxter sigma spectrum infusion system firmware | >= 6.0, <= 6.05 8.0 |
CPE
Remediation
| |
| baxter sigma spectrum infusion system | All versions |