Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Metrics

CVSS 3.x Severity and Vector Strings:

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.

Weakness Enumeration

CWE-IDCWE NameSource
CWE-924Improper Enforcement of Message Integrity During Transmission in a Communication Channel[email protected]