Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
No SSVC data is available for this CVE.
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | CVE | Third Party AdvisoryUS Government Resource |
| http://www.securityfocus.com/bid/105044 | CVE | Third Party AdvisoryVDB Entry |
| https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html | [email protected] | |
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | [email protected] | Third Party AdvisoryUS Government Resource |
| http://www.securityfocus.com/bid/105044 | [email protected] | Third Party AdvisoryVDB Entry |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-287 | Improper Authentication | [email protected] |
| CWE-294 | Authentication Bypass by Capture-replay | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| medtronicdiabetes 508 minimed insulin pump firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 508 minimed insulin pump | All versions |
CPE
Remediation
| |
| medtronicdiabetes 522 paradigm real-time firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 522 paradigm real-time | All versions |
CPE
Remediation
| |
| medtronicdiabetes 722 paradigm real-time firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 722 paradigm real-time | All versions |
CPE
Remediation
| |
| medtronicdiabetes 523 paradigm revel firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 523 paradigm revel | All versions |
CPE
Remediation
| |
| medtronicdiabetes 723 paradigm revel firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 723 paradigm revel | All versions |
CPE
Remediation
| |
| medtronicdiabetes 523k paradigm revel firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 523k paradigm revel | All versions |
CPE
Remediation
| |
| medtronicdiabetes 723k paradigm revel firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 723k paradigm revel | All versions |
CPE
Remediation
| |
| medtronicdiabetes 551 minimed 530g firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 551 minimed 530g | All versions |
CPE
Remediation
| |
| medtronicdiabetes 751 minimed 530g firmware | All versions |
CPE
Remediation
| |
| medtronicdiabetes 751 minimed 530g | All versions |
CPE
Remediation
| |
Change History
8 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| May 22, 2025 | CVE Modified | [email protected] |
| Nov 21, 2024 | CVE Modified | CVE |
| May 14, 2024 | CVE Modified | [email protected] |
| Oct 9, 2019 | CVE Modified | [email protected] |
| Oct 26, 2018 | Initial Analysis | [email protected] |
| Aug 14, 2018 | CVE Modified | [email protected] |
| Aug 14, 2018 | CVE Modified | [email protected] |