Not a U.S. government website. NDD is an independent vulnerability database by Volerion and is not affiliated with or endorsed by NIST or NVD.
CVE-2018-10634 Details
Description
Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
No SSVC data is available for this CVE.
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | CVE | Third Party AdvisoryUS Government Resource |
| http://www.securityfocus.com/bid/105044 | CVE | Third Party AdvisoryVDB Entry |
| https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html | [email protected] | |
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | [email protected] | Third Party AdvisoryUS Government Resource |
| http://www.securityfocus.com/bid/105044 | [email protected] | Third Party AdvisoryVDB Entry |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-319 | Cleartext Transmission of Sensitive Information | [email protected] |
| CWE-319 | Cleartext Transmission of Sensitive Information | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| medtronic minimed paradigm revel mmt-523k firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-523k | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-723k firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-723k | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-723 firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-723 | All versions |
CPE
Remediation
| |
| medtronic minimed 530g mmt-551 firmware | All versions |
CPE
Remediation
| |
| medtronic minimed 530g mmt-551 | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm real-time mmt-522 firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm real-time mmt-522 | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm real-time mmt-722 firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm real-time mmt-722 | All versions |
CPE
Remediation
| |
| medtronic minimed 530g mmt-751 firmware | All versions |
CPE
Remediation
| |
| medtronic minimed 530g mmt-751 | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-523 firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm revel mmt-523 | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm 508 insulin pump firmware | All versions |
CPE
Remediation
| |
| medtronic minimed paradigm 508 insulin pump | All versions |
CPE
Remediation
| |
Change History
9 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| May 22, 2025 | CVE Modified | [email protected] |
| Nov 21, 2024 | CVE Modified | CVE |
| May 14, 2024 | CVE Modified | [email protected] |
| Oct 9, 2019 | CVE Modified | [email protected] |
| Oct 3, 2019 | CWE Remap | [email protected] |
| Oct 31, 2018 | Initial Analysis | [email protected] |
| Aug 14, 2018 | CVE Modified | [email protected] |
| Aug 14, 2018 | CVE Modified | [email protected] |