CVE-2018-10596 Details
Description
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
No SSVC data is available for this CVE.
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01 | CVE | Third Party AdvisoryUS Government Resource |
| https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html | [email protected] | |
| https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01 | [email protected] |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-200 | Exposure of Sensitive Information to an Unauthorized Actor | [email protected] |
| CWE-923 | Improper Restriction of Communication Channel to Intended Endpoints | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| medtronic 2090 carelink programmer firmware | All versions |
CPE
Remediation
| |
| medtronic 2090 carelink programmer | All versions |
CPE
Remediation
| |
Change History
6 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| May 22, 2025 | CVE Modified | [email protected] |
| Nov 21, 2024 | CVE Modified | CVE |
| May 14, 2024 | CVE Modified | [email protected] |
| Oct 9, 2019 | CVE Modified | [email protected] |
| Sep 11, 2018 | Initial Analysis | [email protected] |