CVE-2017-9657 Details
Description
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Metrics
CVSS 4.0 Severity and Vector Strings:
No CVSS 4.0 data is available for this CVE.
CVSS 3.x Severity and Vector Strings:
No data available for CVSS Version 2.0 on this CVE.
No SSVC data is available for this CVE.
References to Advisories, Solutions, and Tools
By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.
| URL | Source(s) | Tag(s) |
|---|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 | CVE | Third Party AdvisoryUS Government Resource |
| https://www.usa.philips.com/healthcare/about/customer-support/product-security | CVE | Vendor Advisory |
| http://www.securityfocus.com/bid/100813 | CVE | Third Party AdvisoryVDB Entry |
| https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 | [email protected] | Third Party AdvisoryUS Government Resource |
| https://www.usa.philips.com/healthcare/about/customer-support/product-security | [email protected] | Vendor Advisory |
| http://www.securityfocus.com/bid/100813 | [email protected] | Third Party AdvisoryVDB Entry |
Weakness Enumeration
| CWE-ID | CWE Name | Source |
|---|---|---|
| CWE-755 | Improper Handling of Exceptional Conditions | [email protected] |
| CWE-460 | Improper Cleanup on Thrown Exception | [email protected] |
Affected Products
| Product | Versions |
|---|---|
| philips intellivue mx40 firmware | < b.06.18 |
CPE
Remediation
| |
| philips intellivue mx40 | All versions |
CPE
Remediation
| |
Change History
7 change records found show changes
| Date | Action | Recorded By |
|---|---|---|
| Jun 17, 2026 | CVE Modified | [email protected] |
| Nov 21, 2024 | CVE Modified | CVE |
| May 14, 2024 | CVE Modified | [email protected] |
| Oct 9, 2019 | CVE Modified | [email protected] |
| Oct 3, 2019 | CWE Remap | [email protected] |
| Jun 12, 2018 | Initial Analysis | [email protected] |
| May 2, 2018 | CVE Modified | [email protected] |