Not a U.S. government website. NDD is an independent vulnerability database by Volerion and is not affiliated with or endorsed by NIST or NVD.
VOLERION
Volerion Security Research

NOT DEFERRED DATABASE

VULNERABILITIES

CVE-2017-12725 Details

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Metrics

CVSS 3.x Severity and Vector Strings:

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.

URLSource(s)Tag(s)
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A CVEThird Party AdvisoryUS Government Resource
http://www.securityfocus.com/bid/100665 CVEThird Party AdvisoryVDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A [email protected]Third Party AdvisoryUS Government Resource
http://www.securityfocus.com/bid/100665 [email protected]Third Party AdvisoryVDB Entry

Weakness Enumeration

CWE-IDCWE NameSource
CWE-798Use of Hard-coded Credentials[email protected]

Affected Products

ProductVersions
smiths-medical medfusion 4000 wireless syringe infusion pump
1.1
1.5
1.6

CPE

  • cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:*
  • cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:*
  • cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:*
  • cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:*

Remediation

  • No remediation found in references.

Change History

5 change records found show changes


QUICK INFO

CVE Dictionary Entry:
CVE-2017-12725
NVD Published Date:
Feb 15, 2018
NVD Last Modified:
Jun 17, 2026
Source:
[email protected]