CVE-2017-12714 Details

Description

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Metrics

CVSS 3.x Severity and Vector Strings:

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving this site. These are references gathered from the official CVE record and are not endorsed by Volerion.

URLSource(s)Tag(s)
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 CVEThird Party AdvisoryUS Government Resource
http://www.securityfocus.com/bid/100523 CVEThird Party AdvisoryVDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 [email protected]Third Party AdvisoryUS Government Resource
http://www.securityfocus.com/bid/100523 [email protected]Third Party AdvisoryVDB Entry

Weakness Enumeration

CWE-IDCWE NameSource
CWE-920Improper Restriction of Power Consumption[email protected]
CWE-920Improper Restriction of Power Consumption[email protected]

Affected Products

ProductVersions

Change History